RK2672AM AC/DC Hipot Tester Manual: Setup, Parameters & Testing
This HTML guide summarizes the key setup, parameter and testing instructions for the RK2672AM AC/DC hipot tester. It is based on the English RK267X series manual that covers the RK2672AM. Use it for quick reference, then consult the original PDF for the complete instructions.
| Model | RK2672AM |
|---|---|
| Test type | AC/DC dielectric withstand (hipot) testing |
| Maximum output voltage | 5kV AC / 5kV DC |
| Maximum test current | 20mA AC / 10mA DC |
| Rated output | 100VA AC at 5.0kV/20mA; 50VA DC at 5.0kV/10mA |
| Test-time setting | 0.0-999 seconds; 0.0 selects continuous testing |
| Input supply | 220VAC +/-10%, 47-63Hz, single phase |
| PLC interface | Optional 9-pin remote terminal |
RK2672AM manual contents
- Safety checks before a hipot test
- AC/DC test setup and connection order
- How to set the test time
- How to set the leakage-current alarm
- Manual and automatic testing procedures
- Optional PLC interface wiring
- RK2672AM setup FAQ
- View or download the original English PDF
Safety checks before a hipot test
- Only trained personnel should operate the tester. Use insulated gloves and an insulating mat where required by your safety procedure.
- Verify the protective ground connection before applying power.
- Keep the tester reset and confirm that the voltage control is at zero before connecting the device under test.
- Connect the current-return lead before the high-voltage lead.
- Do not touch the high-voltage output, test leads or device under test while voltage is present.
- For DC testing, fully discharge the test circuit before replacing or handling the device under test.
AC/DC test setup and connection order
- Turn the voltage adjustment knob fully counterclockwise to zero and keep the instrument in RESET.
- Connect the protective ground and confirm that the selected input supply matches the instrument.
- Connect the current-return lead to the device under test, then connect the high-voltage test lead.
- Select AC or DC test mode, set the test time and set the leakage-current alarm threshold.
- Press START, then slowly turn the voltage control clockwise until the required test voltage is reached.
The high-voltage output remains at zero when the voltage knob is at its minimum position. Always return it to zero before changing connections.
How to set the RK2672AM test time
Set the tester to RESET before changing the timer. Use the time controls to select a value from 0.0 to 999 seconds. A setting of 0.0 runs a continuous manual test; any non-zero value enables a timed test with countdown.
How to set the leakage-current alarm
- Select the required AC or DC mode.
- Use the TEST/PRESET function to display the alarm-current setting.
- Adjust the current preset to the required limit for the test procedure.
- Return the tester to RESET before connecting or changing the device under test.
If measured leakage current exceeds the preset threshold during a test, the RK2672AM stops the high-voltage output and gives an alarm indication.
Manual and automatic testing procedures
Continuous manual test
Set the timer to 0.0, press START and gradually raise the voltage to the required level. Press RESET to end the test. If the measured current exceeds the preset value, the tester alarms and removes the high-voltage output.
Automatic timed test
Choose a non-zero test time, press START and raise the voltage to the required level. The timer counts down. The test completes without an alarm when the countdown reaches zero and the leakage current has remained below the preset threshold.
Optional RK2672AM PLC interface wiring
The optional 9-pin remote terminal accepts dry-contact control and provides voltage-free status contacts. Use momentary contacts for TEST and RESET, and never apply external power to the control inputs.
| Function | Pins |
|---|---|
| Remote TEST input | PIN 1 and PIN 3 |
| Remote RESET input | PIN 1 and PIN 4 |
| Testing status output | PIN 2 and PIN 5 |
| Qualified status output | PIN 8 and PIN 9 |
| Failure status output | PIN 6 and PIN 7 |
The output contacts are rated for a maximum of 12V AC/DC and 100mA. The input is a non-voltage contact; the open-terminal voltage is no more than 10VDC.
RK2672AM setup FAQ
What voltage and current ranges does the RK2672AM provide?
It provides up to 5kV for both AC and DC dielectric withstand testing, with maximum test current of 20mA AC and 10mA DC.
How do I run a continuous hipot test?
Set the test time to 0.0. Press START, raise the voltage gradually, and press RESET when the test is complete.
What does the current preset control?
It sets the leakage-current alarm threshold. The tester stops high-voltage output and alarms if the measured current exceeds that setting.
Can the RK2672AM be controlled by a PLC?
An optional 9-pin PLC interface supports remote TEST and RESET inputs plus testing, qualified and failure status outputs.
RK2672AM product details
For price, package contents, product photos and ordering information, visit the FMF RK2672AM 5kV AC/DC hipot tester product page.
View or download the original RK2672AM English manual
Open the full RK2672AM English manual V1.2 in your browser or download the original PDF.